Salinity
The measurement of salinity is based on the Practical Salinity Scale which uses
the EC measurement. If the user has a standard with known PSU value it may be
used to calibrate the conductivity sensor.
• Select “Salinity” from the “Conductivity calibra-

tion” menu.

• Use the keypad to enter the known salinity value

of the calibration solution. Press to
confirm.

• Fill the calibration beaker with salinity standard of known value.
• Pour additional standard into a second beaker to be used to rinse the sensor.
• Immerse the sensor into the rinse beaker and raise and lower the beaker to

ensure that the EC sensor channels are filled with fresh standard.

• Place the calibration beaker with standard over the EC sensor and dislodge

any trapped gas bubbles. Screw the beaker into place.

• Wait for the reading to stabilize. The stability timer will count down and

firm> will appear.

• Note the temperature and adjust the salinity value if needed.
• Press to save the calibration.
• After confirmation, the following messages are displayed: “Storing” and “Cali-

bration completed”.

• Press to return to the “Calibration” menu.
• Press ESC twice to return to the main menu.
• Press to return to the measurement screen.

Notes

Notes

Notes

Notes

Notes These procedures calibrate the slope value. To calibrate the offset, set

the calibration point at 0 µS/cm and repeat the procedure.
If the temperature input is not within the
acceptable range (0 to 50°C), the mes-
sage “Invalid temperature” is displayed.
If the conductivity input is not within the
acceptable range, the message “Wrong
standard” is displayed.