ISE

From the “GLP” menu, select the “ISE”
option.

• Data regarding the last ISE calibration will be

displayed: standards used, sensor type, time
and date of the calibration.

• Use the arrow keys to scroll through the

stored data for the last 5 calibrations.

• Press ESC to return to the “GLP” menu.
Notes

Notes

Notes

Notes

Notes If no ISE calibration has been performed or if calibration was cleared

using the “Restore factory calib.” option the offset and slope values
are set to default, and the message ”Factory calibration” is displayed.
Press to return to the previous screen.

ORP
• From the “GLP” menu select the “ORP” option.
• Data regarding the last ORP calibration will be

displayed: calibration point, time and date.

• Use the arrow keys to scroll through the stored

data for the last 5 calibrations.

• Press ESC to return to the “GLP” menu.

Notes

Notes

Notes

Notes

Notes If no ORP calibration has been performed or if calibration was cleared

using the “Restore factory calib.” option the offset and slope values
are set to default, and the message ”Factory calibration” is displayed.
Press to return to the previous screen.

Dissolved Oxygen
• From the “GLP” menu select the “Dissolved oxy-

gen” option.

• Data regarding the last D.O. calibration will be

displayed: calibration points, % saturation or
concentration, time and date.

• Use the arrow keys to scroll throught the stored

data for the last 5 calibrations.