Hanna Instruments HI 9829 User Manual

Page 50

background image

50

• Press to accept the calibration point and to continue with second

standard.

• Clean out the calibration beaker and refill with

20.0 FNU standard.

• Immerse the sensor in the 20.0 FNU rinse

beaker and then shake off excess solution.
Place the sensor into the 20.0 FNU calibration
beaker and take measures to eliminate bubbles from the optical area.

• When the reading is stable the display shows the “Ready” message.
• Press to accept the second calibration point. The calibration can

be stopped after the 20FNU point. Press ESC.
Storring will be displayed followed by Calibration
completed. Press or .

Alternately, to continue with third standard:
• Clean out the calibration beaker and refill with

200.0 FNU standard.

• Immerse the sensor in the 200.0 FNU rinse beaker and then shake off excess

solution. Place the sensor into the 200.0 FNU calibration beaker and take
measures to eliminate bubbles from the optical area.

• When the reading is stable the display shows the “Ready” message.
• Press to accept the third point and save the calibration.
• After confirmation, the following messages are displayed: “Storing” and

“Calibration completed”.

• Press to return to the “Calibration” menu.
• To return to the main menu, press ESC twice.
• Press to return to the measurement screen.
• To restore the factory calibration data, select the corresponding option in the

“Turbidity calibration” menu and then press