7.4.1 Preparation
Appendix D – ACCESSORIES lists Hanna solutions used for ORP calibrations.
The calibration should be conducted at temperatures between 20-26°C. The
sensor should be clean and oil free.

7.4.2 Procedure
• From the “Calibration” menu select “Single

param. calibration” and then “ORP calibration”.
The display shows two options: “Custom ORP”
and “Restore factory calib.”.

• For a user calibration select “Custom ORP”.
• Fill a beaker with an ORP test solution (see

APPENDIX D “Accessories”).

• Using the keypad, insert the numerical ORP

value and then press to confirm.

• The stability counter will count down and the

message “Ready” and will be
displayed.

• Press to accept the calibration

point.

• After confirmation, the following messages are displayed: “Storing” and

“Calibration completed”.

• Press OK to return to the Calibration menu.
• Press to return to the measurement screen.
• To restore the factory calibration data, select the corresponding option in the

“ORP calibration” menu and then press