Hanna Instruments HI 9829 User Manual
Page 47

47
• The main display shows the actual reading, while the secondary level displays
the current temperature and the standard value.
• To change the standard value, press
standard values is displayed: 0 µS/cm, 84 µS/cm, 1413 µS/cm, 5.00 mS/cm,
12.88 mS/cm, 80.0 mS/cm and 111.8 mS/cm.
• The third level displays the status message.
• Press
Insert the desired value using the keypad, then press
• When the reading becomes stable, the stability timer will count down and
• After confirmation, the following messages are displayed: “Storing” and “Cali-
bration completed”.
• Press
• Press ESC twice to return to main menu.
• Press
Absolute Conductivity
• Select “Absolute conductivity” from the “Conduc-
tivity calibration” menu.
• Use the keypad to enter the custom value with
the desired resolution. Press
firm.
• Fill the calibration beaker with conductivity stan-
dard with known conductivity at the temperature of standardization.
• Pour additional standard into a second beaker to be used to rinse the sensor.
• Immerse the sensor into the rinse beaker and raise and lower the beaker to
ensure that the EC sensor channels are filled with fresh standard.
• Place the calibration beaker over the EC sensor and dislodge any trapped
bubbles. Screw the beaker into place.
• Wait for the reading to stabilize. The stability timer will count down and firm> will appear. • Note the temperature and adjust the conductivity value if needed. bration completed”. • Press
• Press
• After confirmation, the following messages are displayed: “Storing” and “Cali-
• Press ESC twice to return to the main menu.
• Press