2 probe specifications, Logging probe battery life, Non-logging probe logging probe – Hanna Instruments HI 9829 User Manual
Page 17

17
3.2 PROBE SPECIFICATIONS
Non-logging Probe
Logging Probe
Sample Environment
Fresh, brackish, seawater
Waterproof protection
IP68
Computer Interface
NA
USB PC (HI 76982910)
Internal Battery Type
NA
4 X 1.5V Size AA Alkaline
Typical Battery Life
NA
See below
Memory
NA
140,000 measurements
(single parameter logged)
35,000 measurements
(all parameters logged)
Operating Temperature
-5 to 55° C *
Storage Temperature
-20 to 70° C
Maximum Depth
20 m (66 ft.) *
Dimensions
HI 7609829
HI 7609829
HI 7609829
HI 7609829
HI 7609829 342mm (13.5”),
HI 7629829
HI 7629829
HI 7629829
HI 7629829
HI 7629829 442mm (17.4"),
(without cable)
dia=46 mm (1.8”)
dia=46 mm (1.8")
HI 7619829
HI 7619829
HI 7619829
HI 7619829
HI 7619829 382 mm (15.1”),
HI 7639829
HI 7639829
HI 7639829
HI 7639829
HI 7639829 482 mm (19.0"),
dia=46 mm (1.8”)
dia=46 mm (1.8")
Weight
HI 7609829
HI 7609829
HI 7609829
HI 7609829
HI 7609829 570g (20.1 oz.)
HI 7629829
HI 7629829
HI 7629829
HI 7629829
HI 7629829 775g (27.3 oz.)
(with batteries and sensors)
HI 7619829
HI 7619829
HI 7619829
HI 7619829
HI 7619829 650g (22.9 oz.)
HI 7639829
HI 7639829
HI 7639829
HI 7639829
HI 7639829 819g (28.9oz.)
Cable Specification
Multistrand-multiconductor shielded cable with internal strength
member rated for 68 kg (150 lb) intermittent use
Wetted Materials
Body:
ABS
Threads:
Nylon
Shield:
ABS/ 316 SS
Temp probe: 316 SS
O-rings:
EPDM
* Reduced for ISE sensors
LOGGING PROBE BATTERY LIFE*
Interval
All channels logging
All channels logging
(no averaging)
(10 sample averaging)
1 - 5 sec
72 hours
72 hours
1 min
22 days
11 days
10 min
70 days
65 days
* Continuous logging is contingent on availability of log memory