YSI BODAnalyst User Manual

Page 8

background image

Introduction

Section 1

YSI Incorporated

BOD Analyst for Windows Software

4

Volume of bottles (ml)

300

Dilution Unit

ml

Seed Added to

Bottle

Report Font

Arial

YSI Instrument Parameters Form
Menu:

Setup, YSI Instrument Parameters Form

More Information:

YSI Instrument Parameters Form

Settings Value
Technique

Manual

Stabilize timeout

300 secs.

Sample Hold Time

10 secs.

Sample change (mg/L) 0.05
Blank Hold Time secs. 5
Blank change

0.02

Initial low temp

16

Initial high temp

24

Final low temp

18

Final High temp

22

Supersaturation

9.09

Bottle Numbers

On

Barcode reader (5100 only)

Off

Communications Setup

Menu:

Setup, Communications Setup

More Information:

Communications Settings Form

Settings Value
COM port

Com1:

Baud Rate

9600

Criteria Setup

Menu:

Setup, Criteria

More Information:

BOD Criteria Setup

Settings Value
Maximum Dilution Blank Oxygen Depletion (mg/l)

0.2

Maximum Initial Dissolved Oxygen (mg/l)

9.7

Minimum Final Dissolved Oxygen (mg/l)

1.0

Apply Take 2 leave 1 Rule

On

Minimum (Initial DO - Final DO) (mg/l)

2.0

Remove Depletion caused by Seed

Off

Minimum Oxygen Depletion % for Seed Tests (mg/l)

Disabled

Maximum Oxygen Depletion % for Seed Tests (mg/l)

Disabled

Use Seed that was closest to 50% depletion

On

Minimum Seed Correction Factor (mg/l)

0.6

Maximum Seed Correction Factor (mg/l)

1.0

Minimum Standard Test BOD (mg/l)

163

Maximum Standard Test BOD

237

Check Coefficient of Variation

Off