3.12 MEAS (measurement mode) Key
The MEAS key puts the TI-MVX into it’s primary mode of operation. In
this mode, the user has a complete view of the LCD, as well as control
of the Hot Menu Functions. These hot functions provide the user with
the ability to make crucial adjustments to the display, waveform set-
tings, and grid storage locations without having to search through the menu and
sub menu items. It’s important to first get familiar with moving around in the hot
functions as follows:
Getting Around in the Hot Menus
1) Press the MEAS key at any time to return to
the primarily measurement mode.
2) Press the MEAS key multiple times to tab
right through the menu fields, and the ESC key
multiple times to tab left through the menu fields,
until the desired hot function is highlighted.
Now that you’re familiar with activating and moving amongst the hot function
fields, let’s have a look at how to adjust or change the values of these fields:
Adjusting the Values of the Hot Menus
1) Use the UP, DOWN, LEFT, and RIGHT arrow keys to increase and decrease
the values of the hot function fields.
2) Repeat step 1 until the desired value has been achieved.
Alternatively, the delay, width, gain, and threshold hot function fields can be
changed using the Digit Edit Box as follows:
1) Press the ENTER key to display the Digits Edit Box.
2) Press the UP and DOWN arrow keys to scroll the highlighted value.
3) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
4) Repeat steps 2 & 3 until the delay, width, gain, or threshold number is correct-
ly displayed.
5) Press the OK key to set and return to the measure screen, or ESC to cancel
entering the delay, width, gain, or threshold value.
-18-
4) Press the MEAS key to return to the measurement screen and begin taking
readings.
Important: Be sure to do a Probe Zero after change the polarity!
9.6
Rectified Waveform (Filled/unfilled)
The TI-MVX has yet another feature primarily included for user personal viewing
preference. When the TI-MVX is in the Rectified Display View, the waveform can
be completely filled in black, or simply outlined. The procedures to toggle this
feature, are outlined below:
Toggle Rect Wave (filled / Unfilled)
Note: Before toggling the Rect Wave, the TI-MVX should be set to the RECT
display view option. Please refer to page 45 for information on selecting the
Display Views.
1) Press the MENU key once to activate
the menu items tab. Press the MENU key
multiple times to tab right, and the ESC
key multiple times to tab left, until the
TUNE menu is highlighted and displaying
the submenu items.
2) Use the UP and DOWN arrow keys to
scroll through the sub menu items until RECT WAVEFORM is highlighted.
3) Use the LEFT and RIGHT arrow keys to toggle the RECT WAVEFORM filled
or unfilled.
4) Press the MEAS key to return to the measurement screen and begin taking
readings.
-63-