Power Requirements
July 2014
TMX428 Owner’s Manual
3-1
Power Requirements
The TRACKMASTER
®
is designed to operate on a dedicated 110-120 VAC 20-amp power
supply or optional 220/240 VAC 15-amp. Make sure that the treadmill is connected to an
outlet that looks like the following illustration.
This product is equipped with a three-wire grounding-type plug. The plug will only fit into a
grounding-type outlet. This safety feature must not be disabled. Contact a qualified
electrician if you are unable to insert the plug into your outlet, or uncertain if the outlet meets
local electrical codes. Polarized outlets such as NEMA 5-20 and CEE7/7 must be verified for
proper polarity configuration before plugging in the TRACKMASTER
®
. Incorrect
polarization of the outlet could cause failure of onboard electrical components or cause
electrical shock. Proper grounding is necessary for the equipment to meet acceptable current
leakage standards consistent with the standards to which it was certified.
TRACKMASTER
®
treadmills must be grounded to reduce the risk of
electrical shock. If a malfunction occurs, grounding provides a path of least
resistance for an electric current. Ungrounded connections must not be used.
No other equipment may be used on the electrical circuit with the
TRACKMASTER
®
. Do not use extension cords. Using a shared or
unreliable circuit can also cause the TRACKMASTER
®
to unexpectedly
shut off, potentially resulting in injury to the patient.
Ensure the master power switch is in the off position before plugging
in the TRACKMASTER
®.
A power surge could damage the
sophisticated electronic system of the treadmill.
3
110-120VAC
220/240 VAC
220/240 VAC
230 VAC
NEMA 5-20R
NEMA 6-15R
CEE 7/7 EURO
UK BS1363
Single Phase
Split Phase
Option
Option