beautypg.com

15 specifications, Specifications, Cardioperfect workstation – Welch Allyn CardioPerfect Workstation SpiroPerfect Module - User Manual User Manual

Page 79: Spiroperfect module – user manual

background image

CardioPerfect Workstation

SpiroPerfect Module

– User Manual

DIR 80012334 Ver. F

79 / 95

15 Specifications

Specification

Description

SpiroPerfect

Computer based full diagnostic spirometer

Tests

FVC, SVC, MVV, Pre-Post BD

Sensor Type

Pneumotach

Power Equipment

None, obtained from USB port

Accuracy

Meets or exceeds ATS/ERS 2005 standard

Reproducibility

Meets or exceeds ATS/ERS 2005 standard

Volume Range

0-14 L

Flow Range

+- 14 L/sec

Predictive Norms

For Predictive Norms included, see 12.2

Additional predictive norms can be added upon customer request

Interpretation

1991 ATS Interpretation Standards.
Automatic interpretation can be disabled.
Manual Interpretation available.
Lung Age calculation

Reports

FVC - Volume / Time
FVC - Flow / Volume
FVC - Both

– Volume / Time and Flow / Volume

SVC

– Volume / Time

Incentive graphic

Fireman

Parameters

FVC, FIVC, FIV1, FIV1%, FEV0.5, FEV1, FEV2, FEV3, FEV5, FEV6, FEV0.5,
FEV0.5%, FEV1%, FEV1/FVC, FEV2%, FEV3%, FEV5%, FEV6%, PEF, FEF25,
FEF50, FEF75, FEF0.2-1.2, FEF25-75, FEF75-85, PIF, FIF50, FEF50/FIF50,
FEV1/FEV6, FET, MEF25, MEF50, MEF75
SVC, ERV, IRV, VT, IC, BF, MV, Tin, Tex, Tin/Tex
MVV, MV, VT, BF, DFRC

Quality Checks

ATS Acceptability and ATS Reproducibility checks
Audio and visual incentive for assistance in coaching patients

Connectivity

Compatible with CardioPerfect Workstation
Export compatible with most electronic medical records programs
Available in multi user network
Telemedicine option for e-mail transfer

Storage and

Environment

Temperatures between

–20 °C (-4 °F) and 50 °C (122 °F).

Relative humidity between 15 and 95% (non-condensing).

Atmospheric Pressure of 500 hPa (mbar) to 1,060 hPa (mbar).