15 specifications, Specifications, Cardioperfect workstation – Welch Allyn CardioPerfect Workstation SpiroPerfect Module - User Manual User Manual
Page 79: Spiroperfect module – user manual

CardioPerfect Workstation
SpiroPerfect Module
– User Manual
DIR 80012334 Ver. F
79 / 95
15 Specifications
Specification
Description
SpiroPerfect
Computer based full diagnostic spirometer
Tests
FVC, SVC, MVV, Pre-Post BD
Sensor Type
Pneumotach
Power Equipment
None, obtained from USB port
Accuracy
Meets or exceeds ATS/ERS 2005 standard
Reproducibility
Meets or exceeds ATS/ERS 2005 standard
Volume Range
0-14 L
Flow Range
+- 14 L/sec
Predictive Norms
For Predictive Norms included, see 12.2
Additional predictive norms can be added upon customer request
Interpretation
1991 ATS Interpretation Standards.
Automatic interpretation can be disabled.
Manual Interpretation available.
Lung Age calculation
Reports
FVC - Volume / Time
FVC - Flow / Volume
FVC - Both
– Volume / Time and Flow / Volume
SVC
– Volume / Time
Incentive graphic
Fireman
Parameters
FVC, FIVC, FIV1, FIV1%, FEV0.5, FEV1, FEV2, FEV3, FEV5, FEV6, FEV0.5,
FEV0.5%, FEV1%, FEV1/FVC, FEV2%, FEV3%, FEV5%, FEV6%, PEF, FEF25,
FEF50, FEF75, FEF0.2-1.2, FEF25-75, FEF75-85, PIF, FIF50, FEF50/FIF50,
FEV1/FEV6, FET, MEF25, MEF50, MEF75
SVC, ERV, IRV, VT, IC, BF, MV, Tin, Tex, Tin/Tex
MVV, MV, VT, BF, DFRC
Quality Checks
ATS Acceptability and ATS Reproducibility checks
Audio and visual incentive for assistance in coaching patients
Connectivity
Compatible with CardioPerfect Workstation
Export compatible with most electronic medical records programs
Available in multi user network
Telemedicine option for e-mail transfer
Storage and
Environment
Temperatures between
–20 °C (-4 °F) and 50 °C (122 °F).
Relative humidity between 15 and 95% (non-condensing).
Atmospheric Pressure of 500 hPa (mbar) to 1,060 hPa (mbar).