3 cleaning the calibration syringe, 4 ordering information for replacement parts, Cleaning the calibration syringe – Welch Allyn CardioPerfect Workstation SpiroPerfect Module - User Manual User Manual
Page 75: Ordering information for replacement parts, Cardioperfect workstation, Spiroperfect module – user manual, Warning, Caution

CardioPerfect Workstation
SpiroPerfect Module
– User Manual
DIR 80012334 Ver. F
75 / 95
If you choose to clean the calibrations syringe, clean the outer
surface of the syringe with only the following solutions or wipes:
Solution of dish soap and water, ½ tsp per cup of water
Solution of bleach and water, 1 part bleach (6% sodium
hypochlorite) with 9 parts water
Isopropyl Alcohol and water, 70% by volume
PDI Sani-Cloth Plus wipes (14.85% Isopropanol)
Cavi-Wipes (17.2% Isopropanol)
Warning
Satisfactory maintenance procedures must be implemented, or
equipment failure and health hazards may result. Only qualified
service personnel should repair the equipment. See “Limited
Warranty” and “Service Policy”, page 4.
To prevent cross-contamination, do not try to clean the flow
transducers and nose clips. Discard these items after a single
patient use. Wear gloves when replacing flow transducers, and
wash hands after touching them.
Caution
Do not clean the pressure tubing or sensor. Trapped moisture
could affect accuracy.
Replace the pressure tubing when it becomes dirty. Recalibrate
after replacement.
Replace the sensor when it becomes faulty. See section 13.4
Ordering Information for Replacement Parts on page 75.
13.3
Cleaning the Calibration Syringe
Clean the outer surface of the syringe with any of the following solutions or wipes:
Solution of dish soap and water, ½ tsp per cup of water
Solution of bleach and water, 1 part bleach (6% sodium hypochlorite) with 9 parts water
Isopropyl Alcohol and water, 70% by volume
PDI Sani-Cloth Plus wipes (14.85% Isopropanol)
Cavi-Wipes (17.2% Isopropanol)
13.4
Ordering Information for Replacement Parts
The following parts must be replaced as noted:
flow transducers & nose clips
– Replace for each new patient.
pressure tubing
– Replace when dirty.
sensor
– Replace when faulty.
To order parts, call the Welch Allyn Technical Support Center.
Warning
Discard all spirometry components according to local regulations.
Use of components other than those recommended by Welch Allyn may compromise product
performance. The Welch Allyn warranty can only be honored if you use Welch Allyn approved