7.3

View Calibration Results

After calibration the Verify Calibration Results window will appear.

Figure 7.6 Verify Calibration Results




Each row gives the result of a stroke, the last row gives the Averages. The following columns
are given:
L/s

Results per stroke:
The speed in the L/s plunger should be moved, as indicated by the blue
calibration bar during the strokes.

Exp. Vol

The expiration volume reached by pushing in the plunger.

Exp.%

The expiration deviation from the actual volume in percent.

Insp. Vol

The inspiration volume reached by pulling out the plunger.

Insp.%

The inspiration deviation percentage

Avg.Vol

The averages of inspiration volume and expiration volume.

Abs.Avg.% The absolute average deviation percentage between inspiration volume and

expiration volume.


The Averages row gives the averages reached for all strokes. Please refer to this row for
improving the calibration results.

After you have accepted the results you can view these values in the table below the graph by
selecting the Calibrate Results button.

Note:
Abs.Avg% should be:

-

<1% to meet the social
security standard

-

<3% to meet ATS standard

(|Exp.%| + |Insp.%|)/2=Abs.Avg.%