Final Result
A patient’s best effort is a measurement calculated from a set of efforts. To
determine the method in which best effort is calculated, choose from these
options:
Best effort
Defines best effort as the single best effort in a set of
efforts per effort type (best FVC-pre, best FVC post,
best SVC). This ATS-recommended method uses the
effort with the highest sum of FVC + FEV1, or the effort
with the highest SVC value. (For details, see the
document noted in reference 5.)
Best Composite
Defines best effort as a composite of the highest
parameter values across all selected efforts.
Calibration
syringe
Default value for the volume of the calibration syringe. Select the Syringe
Volume from the list.
Calibration
reminder
Check this box to receive a calibration reminder pop-up daily, weekly or
monthly.
Pressure
Determines the unit of Pressure. Check the preferred unit.
Flow
Determines the unit of Flow on the axis of the graph, possible options are
L/s or L/m.
Temperature
Determines the unit of Temperature, possible options are °C or °F.
5.2
Viewing Tab
To display the Viewing tab:
1. Choose File
2. Select Settings > Spirometry > Viewing
The following screen appears: