Setting
Description
Calibration
Error Tolerance
Select the measured calibration accuracy to be within 1% or 3% of the
syringe volume.
Tip: See page 28 on how to change the default setting.
Temperature
The ambient temperature
See Operating Environment Specifications on page 80
Humidity
The ambient humidity
See Operating Environment Specifications on page 80
Pressure
The ambient pressure
See Operating Environment Specifications on page 80
WARNING
For the ambient settings pressure field please enter the
pressure as given by a barometer in the immediate vicinity.
Do not enter the normalized sea-level pressure as commonly
listed on internet sites on meteorological data resources.
2. Press the Next button to continue.
The following screen appears depending on your setup:
Figure 7.2 Calibration window
Beneath the Calibration window the calibration instruction and message window is
displayed, giving instruction for the calibration procedure.
Note: Please check the calibration and lot code if you cannot continue to the
Calibration window.
3. Connect the syringe to the new flow transducer.
4. Fill the syringe by pulling the plunger completely out.
5. Press the Calibrate button in the window.
6.
Wait until the messages ‘Initializing sensor. Opening sensor, please wait...’
disappears.