beautypg.com

Cardioperfect workstation, Spiroperfect module – user manual, Warning – Welch Allyn CardioPerfect Workstation SpiroPerfect Module - User Manual User Manual

Page 25

background image

CardioPerfect Workstation

SpiroPerfect Module

– User Manual

DIR 80012334 Ver. F

25 / 95

Figure 5.4 Spirometry Parameters tab


Select parameters for three categories:
Setting

Description

Measurements

Parameters selected in the Measurements column are displayed in the
Measurements Tab of the SpiroPerfect module.

Six parameters

Parameters selected in the Six parameters column are displayed in the six
parameters table of the module’s Parameter area. A maximum of six
parameters can be selected per test type. For FVC, a minimum of three
parameters is required.

Printer

Parameters selected in the Printer column are printed on the reports.

WARNING

Selecting more than 15 parameters for
printing may result in the list of printed
parameters being truncated on all but the
Best FVC report.


Parameters Measured
FVC testing

FVC
FEV2
FEV1%
PEF
FEF75-85

FIVC
FEV3
FEV1/FVC
FEF25
PIF

FIV1
FEV5
FEV2%
FEF50
FIF50

FIV1%
FEV6
FEV3%
FEF75
FEF50/FIF50

FEV0.5
FIV0.5
FEV5%
FEF0.2-1.2
FEV1/FEV6

FEV1
FEV0.5%
FEV6%
FEF25-75
FET