beautypg.com

Nibp settings – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual

Page 70

background image

66

Monitoring and Measurements

Welch Allyn 1500 Patient Monitor

NIBP settings

The parameter settings are selected with the trim knob, described previously (see

“Settings via a parameter field”

on page 31)

.

The default settings are in bold.

Parameter

Description

NIBP Interval

Off

, 3, 5, 10, 15, 30, 60

Format

Sys/Dia

or Mean. This defines the main measurement to be

displayed, that is the larger measurement displayed in the NIBP box.
The secondary measurement is displayed smaller by the side of the
defined measurement. Note that the mean measurement is shown in
brackets.

Initial Inflation Pressure

Adult: 160
Pediatric: 120
Neonate: 90

NIBPs (systolic blood pressure)
upper/lower limit

Range: 30 - 300
Adult 220/75
Pediatric: 145/75
Neonate: 100/50

NIBPd (diastolic blood pressure)
upper/lower limit

Range: 20 - 235
Adult 110/35
Pediatric: 100/35
Neonate: 70/30

NIBPm (mean pressure) upper/
lower limit

Range: 20 - 255
Adult 120/50
Pediatric: 110/50
Neonate: 80/35