Maintenance

Maintenance interval

This software controlled monitor has undergone a software risk analysis to minimize any
hazards associated with software defects.

The regular system maintenance must include a functional test according to the
manufacturer's instructions. The test results should be recorded (see

“Inspection and

checklist report”

on page 104).

Maintenance work not described in this section, e.g. battery replacement, may only be
accomplished by a qualified technician.

The following table indicates the intervals and responsibilities of the maintenance work
required. Local regulations in your country may stipulate additional or different inspection
intervals and tests.

Interval

Maintenance

Responsible

Before use

Visual inspection of the monitor
and cables

User

Every 6 months

Visual inspection of the monitor
and cables

User

Button check

Speaker check

LED check

Alarm check

Every 12 months

Yearly test and test after repair
according to IEC/EN 62353.

Qualified technician

CO

2

Calibration

1

1.

The need for calibration is based upon physical component changes that occur during use. The module requires
its first calibration after 1200 operating hours or one calendar year, whichever comes sooner, and then after each
4000 operating hours or once a year, whichever comes sooner. The message Calibration Due appears when the
hourly limit is reached. It is advisable to calibrate in the one-year maintenance program especially if the monitor is
used for intermittent, short term use typical of patient monitors.