Directions for use

Settings

119

The Administrator menu is as follows:

Sub menu

Parameter

Description

Configuration

Monitor data such as serial number, software version etc. This
is provided for information only and no settings can be made.

Communications

This provides the communication settings for service
personnel only.

Alarms
(password
protected - see
above)

Alarm Silence Time

1, 1.5 or 2 minutes. Time for which an audible alarm is
silenced.

Alarm Suspend Time

1, 1.5 or 2 minutes. Time for which all alarms are suspended.

Can disable HR / PR
Alarms

Enabled/

Disabled. The menu item can prevent users from

being able to turn off HR or PR alarms. When the setting is
"Disabled", a user cannot turn off the HR or PR alarms.

Alarm Delay

On/

Off. If the alarm validation is enabled, alarm limits must be

exceeded for at least 6 seconds for an alarm to be issued.

Second Speaker Time

0 to 3 minutes (2 minutes). Time after which the secondary
speaker is enabled. For lethal and high level alarms the
secondary speaker is activated in 30 seconds.

Audio Off

Yes, or No. Disables the audible alarm indefinitely until reset
manually, or when a new patient is defined, or the monitor
switched off.

System
(password
protected - see
previous page)

Height units

Select Inches (in) or centimeters (cm)- sets the monitor’s
measurement units.

Weight units

Select pounds (lbs) or kilograms (kg)- sets the monitor’s
measurement units. In neonate mode the units are
automatically switched to grams.

Acuity Enabled

Yes/No. Connects to Acuity. Note that Acuity can only be
enabled when Enable Acuity connection in the service setup is
set.

Set Date and Time

Entry of the year/month/day/hour/minute.

Save User Defaults

With this function, values changed by the user are saved.

Show Event Log

Display of the monitor event log (see

“Event log screen and

CO2 calibration”

on page 120).

Service

This provides settings, options and service information for
service personnel only.

Factory

This menu is for factory use only.