108

Maintenance

Welch Allyn 1500 Patient Monitor

EMC compliance

The monitor is designed for use in an electromagnetic environment in accordance with
IEC/EN 60601-1-2, tables 201, 202 and 204. If the monitor is used in the vicinity of
equipment labelled with the symbol “Non-ionic electromagnetic radiation"

, check the

recommended minimum distance according to IEC/EN 60101-1-2, table 206. For further
details, please refer to the service manual.

The following table lists devices and their typical frequency ranges and transmitting
power, and the resulting minimum distances.

For transmitters not included in the above table, the recommended distance (d in meters)
can be calculated using the following formulas:

Frequency range 0.15 – 80 MHz

d= 3.5 ÷ 3V x √P

Frequency range 80 – 800 MHz

d = 3.5 ÷ 3V/m x √P

Frequency range 800 MHz – 2.5 GHz

d = 7 ÷ 3V/m x √P

d = recommended distance in meters

P = transmitting power in watts

V = volts

m= meters

HF source

Transmitter

frequency

[MHz]

Power

P

[W]

Distance

d

[m]

Radio telephone (micro cellular) CT1+, CT2, CT3

885-887

0.010

0.23

Cordless DECT telephone, WLAN, UMTS handy

1880-2500

0.25

1.17

Mobile phone, handy USA

850/1900

0.6

1.8

Mobile phone, handy

GSM900 900

2

3.3

GSM850, NMT900, DCS 1800

850, 900, 1,800

1

2.3

Walkie-talkie (rescue service, police, fire
brigade, service)

81-470

5

2.6

Mobile telephone system (rescue service, police,
fire brigade)

81-470

100

11.7