Arrhythmia

Arrhythmia settings

1.

Press the Setup button

.

2. Select Arrhythmia.

The default settings are in bold.

Note

The full arrhythmia menu is only displayed when the full arrhythmia option is
licensed. When the full arrhythmia option has not been enabled, settings are
made for VFib, Asystol and Vtach only.

Note

The arrhythmia menu is not available when the monitor is connected to Acuity.

Main menu

Parameter

Description

Setup Arrhythmia

VFib

Alarm Level - Lethal (cannot be changed), Print
on alarm Yes/No.

Asystole

Alarm Level - Lethal (cannot be changed), Print
on alarm Yes/No.

VTach

VTach Limit 100 to 200 (125), Alarm Level - Lethal
(cannot be changed), Print on alarm Yes/No.

PVC run

PVC run 3 to 6 (6) Alarm Level (High, Medium,
Low, Off), Print on alarm Yes/No.

VRyhthm

Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.

Couplet

Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.

PVC/min

PVC/min 1to 30 (30) Alarm Level (High, Medium,
Low, Off), Print on alarm Yes/No.

Bigeminy

Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.