Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual
Page 55
Directions for use
Monitoring and Measurements
51
Main menu
Parameter
Description
Setup Arrhythmia
(continued)
Trigeminy
Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.
Tachycardia
Tachycardia 150 to 250 (180) Alarm Level (High,
Medium, Low, Off), Print on alarm Yes/No.
Bradycardia
Bradycardia 20 to 100 (45) Alarm Level (High,
Medium, Low, Off), Print on alarm Yes/No.
Pause
Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.
Irregular (irregular rhythm)
Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No. This is an irregularity in the R to R
interval over a series of at least 16 non-
ventricular beats. The number of beats analyzed
is given in options.
Non-capture (pacemaker
non-capture)
Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No. This is for pacemaker patients
with the analyze pacers option enabled (see
options below) - a beat does not directly follow a
pacer.
Options
Irregular Limit
Yes/No, set limit between 45 and 120 (45)
Single ECG
Yes/No, select lead
I
, II, III, or V. Select this
option to analyze one ECG lead.
Pacer Display
Yes/No displays pacer pulses relative to time but
not representative of either amplitude, duration
or polarity.
Analyze Pacers
Yes/No enables non-capture (see above).
Arrhythmia OFF
Sets all alarms to off except VFib, Asystole and
VTach which remain set at the highest alarm
level.
Presets
Resets all arrhythmia settings to the default.
ST / ARR Relearn
Approximately 15 - 20 complexes are used to set
the parameters (duration, amplitude, etc.) for
Arrhythmia analysis. Select this option to
redefine the template used.