Directions for use

Monitoring and Measurements

51

Main menu

Parameter

Description

Setup Arrhythmia
(continued)

Trigeminy

Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.

Tachycardia

Tachycardia 150 to 250 (180) Alarm Level (High,
Medium, Low, Off), Print on alarm Yes/No.

Bradycardia

Bradycardia 20 to 100 (45) Alarm Level (High,
Medium, Low, Off), Print on alarm Yes/No.

Pause

Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.

Irregular (irregular rhythm)

Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No. This is an irregularity in the R to R
interval over a series of at least 16 non-
ventricular beats. The number of beats analyzed
is given in options.

Non-capture (pacemaker
non-capture)

Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No. This is for pacemaker patients
with the analyze pacers option enabled (see
options below) - a beat does not directly follow a
pacer.

Options

Irregular Limit

Yes/No, set limit between 45 and 120 (45)

Single ECG

Yes/No, select lead

I

, II, III, or V. Select this

option to analyze one ECG lead.

Pacer Display

Yes/No displays pacer pulses relative to time but
not representative of either amplitude, duration
or polarity.

Analyze Pacers

Yes/No enables non-capture (see above).

Arrhythmia OFF

Sets all alarms to off except VFib, Asystole and
VTach which remain set at the highest alarm
level.

Presets

Resets all arrhythmia settings to the default.

ST / ARR Relearn

Approximately 15 - 20 complexes are used to set
the parameters (duration, amplitude, etc.) for
Arrhythmia analysis. Select this option to
redefine the template used.