Respiration, Temperature, Respiration temperature – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual
Page 136

132
Technical data
Welch Allyn 1500 Patient Monitor
Respiration
Temperature
ECG amplifier
Sampling frequency
1000 Hz
Pacemaker detection ± 2 to ± 700 mV / 0.1 to 2 ms
Pacemaker rejection ± 2 to ± 700 mV / 0.1 to 2 ms
Note: Pacemaker signals can differ from one pacemaker to the next. Rate
meters may continue to count the pacemaker rate during occurrences of
cardiac arrest or some arrhythmias, mainly with pacemakers generating
high amplitudes ( > 20 mV) or those generating overshoot. Pacemaker
patients should be kept under close or constant observation.
Protection
Fully isolated, defibrillation protected >5 kV
Line frequency filter
50 or 60 Hz sinusoidal interferences filtered by means of adaptive digital
filtering.
Respiration rate range
0 to 200 breaths / min (pediatric: 0 to 120 breaths per minute)
Connector
Shared with ECG
Signal
28 kHz square wave ± 2.5 V
Patient current
max. 80 :A
Dynamic impedance
range
1 k to1.5 kS, variation of 0.1 to 3 S
Sampling Rate
250 Hz
Respiration rate accuracy ± 1 digit
RR display update interval Max. 2 s
Channels
One or two channels
Sensor
YSI 401, rectal, skin or ear
Amplifier
Fully isolated, defibrillation protected >5kV
Sampling Frequency
125 Hz
Measurement interval
1x per second
Measurement range
15 °C to 45 °C (59 °F to 113°F)
Resolution
0.1 °C (0.1 °F)
Accuracy
+ 0.1° C (+ 0.1° F)