80 RL-TS Series Operation Manual
Message
Cause
Remedy
The balance is influenced by static 
electricity or noise. 
- Unplug the AC adapter from the receptacle and then
turn the power on again.
- If this error occurs again, electric components may
have failed. Contact our Sales Office or Technical 
Service Division. 
The balance is influenced by static 
electricity or noise. 
- Unplug the AC adapter from the receptacle and then
turn the power on again.
- If this error occurs again, electric components may
have failed. Contact our Sales Office or Technical 
Service Division. 
The weight of a sample is too light at a 
sampling during parts counting, or reference 
weight saving during percentage weighing. 
Use a heavier sample by referring to the Specifications 
(P. 
) to check the minimum unit weight and the
percentage weighing weight limit.
- At addition operation, you placed
additional samples on twice.
- At addition operation, you unloaded some
samples or you pressed the key without 
adding samples. 
- After setting the display to “0” (by unloading the
previous sample), place the next sample to continue 
addition operation. 
- Addition operation is impossible when 0 or a
negative value is displayed. Place a sample to 
continue addition operation. 
No inputs are sent from the weight sensor.
- Unplug the AC adapter from the receptacle and then
turn the power on again.
- If this error occurs again, the sensor may have failed.
Contact our Sales Office or Technical Service 
Division. 
Because the balance is unstable, 
initialization cannot be completed. 
The balance may be affected by an external influence 
such as wind and vibration. Relocate the balance by 
referring to the section “Do not use the balance in the 
following places” (P. v). 
33B