12b2.5 function setting basics, 12b 2.5, Function setting basics – Rice Lake TS Series, Rice Lake Weighing Systems - Operation Manual User Manual
Page 26: 5 function setting basics, Reference

22 RL-TS Series Operation Manual
12B
2.5 Function Setting Basics
To set the settings of the balance, use the function keys.
This section describes the basic operations of function setting
Pressing the [Print] key when a function is being set can cancel the setting
and return to measurement mode.
1
Set to the function setting mode.
After display changes,
release the finger.
Hold down the [Function] key and release the
finger after display has changed to “Func.”
The first function item is displayed.
Keeping the [Function] key held down will
switch the balance to another mode. If this
happens, press the [Print] key to cancel the
setting and redo the setting from the beginning.
2
Select a setting item.
By pressing the [Function] key, select an item to
be set.
Example: By pressing the [Function] key once,
select “2.SEL” (additional function). The first
setting “2.SEL 0” is displayed.
3
Select a setting.
Pressing the [Zero/Tare] key selects a setting.
Pressing the key toggles the settings in turn.
After the last setting is displayed, the next
setting displayed is the first one.
Example: By pressing the [Zero/Tare] key twice,
select “2.SEL 2.”
4
Save the setting.
Press the [Target] key to complete the setting.
The display returns to showing the sample’s
normal weight.
To cancel, press the [Print] key.
Reference