Rice Lake TS Series, Rice Lake Weighing Systems - Operation Manual User Manual
Page 79

RL-TS Series Operation Manual 75
Command format
(1) Tare range (zero-point adjustment) command
C1
C2
Code (C1) Code (C2)
Description
Value
Response
T
(SP) 54H
20H
- Tare range setting
- Zero-point
adjustment
None
A00: Successful completion
E01: Command error
E04: A tare range (zero-point
adjustment) cannot be set
(Range violation, weight error,
etc.).
(2) Set output control command
C1
C2
Code (C1) Code (C2)
Description
Response
O
0
4FH
-{}-30H
Stop output.
A00: Successful
completion
E01: Command
error
O
1
4FH
31H
Output continuously at all times.
O
2
4FH
32H
Output continuously if stable (Stop output if
unstable).
O
3
4FH
33H
Output once when the [Print] key is pressed
(whether the balance is stable or unstable).
O
4
4FH
34H
Output once when the balance is stable. Output
when a sample is lifted to cause the display to
indicate a value below zero, and then another
sample is placed to make the balance stable.
O
5
4FH
35H
Output once when the balance is stable. Stop
output when unstable. Output once when the
balance is stabilized again (the output includes
zero) even if it is not reloaded.
O
6
4FH
36H
Output once when the balance is stable. Output
continuously when unstable. Output is stopped
after a single output when the balance is stable
even if it is not reloaded.
O
7
4FH
37H
Output once when the [Print] key is pressed if
the balance is stable.
O
8
4FH
38H
Output once immediately.
O
9
4FH
39H
Output once after stabilized.
- Commands O0 to O7 have the same workings as the output control set by the function setting.
- Commands O8 and O9 are used to request data from the balance.
- Once executed, O0 to O7 commands are held. However, the status is reset to the function setting when the
balance is turned on again.