32b7.1 error messages, 32b 7.1, Error messages – Rice Lake TS Series, Rice Lake Weighing Systems - Operation Manual User Manual
Page 83: 1 error messages

RL-TS Series Operation Manual 79
32B
7.1 Error Messages
Message
Cause
Remedy
- The weight of the sample is over the
weighing capacity.
- The number of digits in the addition result
or calculation result went over the number
that can be displayed.
- Unload the sample to weigh it in some portioned-out
measurements.
- Replace the tare with a lighter one.
- If the error message does not disappear even when
nothing is placed on the pan, mechanical parts may
have failed. Contact our Sales Office or Technical
Service Division.
- First, clear the addition result. Then execute addition
again.
- The coefficient used in unit converting is too small.
Set a greater coefficient.
The minus-value load exceeded the lower
limit.
- The pan or the pan base may not be properly set.
Check them, paying attention to whether they are in
contact with an external object.
- If the error message does not disappear even if the
pan and the pan base are properly set, mechanical
parts may have failed. Contact our Sales Office or
Technical Service Division.
The reference weight used during span
adjustment with an external weight is far
less than 50% of the weighing capacity.
For span adjustment with an external weight, use a
weight that weighs as close to the weighing capacity as
possible.
An error over 1.0% was detected in span
adjustment with an external weight, or the
balance failed.
For span adjustment with an external weight, check
that a correct weight is placed and that no objects other
than the weight are placed. Then, execute span
adjustment again.