28b5.6 improving the stability of the balance, 28b 5.6 improving the stability of the balance, 6 improving the stability of – Rice Lake TS Series, Rice Lake Weighing Systems - Operation Manual User Manual
Page 66: The balance, 6 improving the stability of the balance, Reference

62 RL-TS Series Operation Manual
28B
5.6 Improving the Stability of the Balance
When the balance is stable, “” is lit in the upper left of the display.
When displayed values flicker and stabilized display blinks, it indicates that the balance is influenced by
wind or vibration. In these situations, making a setting change can improve stability.
As greater values are set in the function setting of “stability judgment (4.5.d),” “response speed
(5.Re.),” and “minimum readability setting (b2.d.A, b4.d.b),” stability will be improved more.
Relationship between each function setting and wind/vibration influences
Wind/vibration
influences
Stability judgment
Response speed
Minimum readability
setting
Small
Big
1
1
1
2
2
2
3
3
3
4
4
4
5
5
In each of the functions, if wind and vibration influences are small, select 1 or 2. Set 3 to 5 for great
influences.
1
Set to the function setting mode.
(Refer to “Section 2-5: Function Setting
).)
Press and hold the [Function] key. After “Func” is
displayed, release the finger.
2
Select each function.
Press the [Function] key several times to select the functions
(See the above table).
4.5.d. =
Stability judgment
5.r.E.=
Response speed
b2.d.A =
Minimum readability setting (A)
b4.d.b =
Minimum readability setting (B)
3
Select a setting.
Press the [Zero/Tare] key to select the setting values of each
function (see the above table).
Reference