Analog output setting (analog models only) – Rice Lake Signal Conditioning Transmitter/Indicators User Manual
Page 53

Operation 49
At the time of writing, the setpoints, hysteresis values are saved to the RAM and will be lost upon the next
power-off; to store them permanently to the EEPROM so that they are maintained at power-on, the ‘99’
command of the Command Register must be sent.
Weight (Span) Calibration Commands (With Test Weights)
The instrument calibration can be changed via MODBUS. To carry out this procedure, the system must be
unloaded and the weight value display reset to zero with the command ‘100’ of the Command Register. Then, a
load must be placed on the system and the correct weight value must be sent to the registers 40037-40038; to save
this value, send the control ‘101’ from the Command Register. If the operation is successfully completed, the two
test weight registers are set to zero.
Analog Output Setting (Analog Models Only)
Write the weight in the registers “Weight value corresponding to the Full Scale of analog output H” (40045) and
“Weight value corresponding to the Full Scale of analog output L” (40046) or write the weight in the registers
“weight value corresponding to the ZERO of the analog output H” (40043) and “weight value corresponding to the
ZERO of the analog output L” (40044). After writing the value, send the command 99 from the Command Register
to save it to EEPROM memory.
40017
SETPOINT 1 H
Only after command ‘99’ of the
COMMAND REGISTER
R/W
40018
SETPOINT 1 L
R/W
40019
SETPOINT 2 H
R/W
40020
SETPOINT 2 L
R/W
40021
SETPOINT 3 H
R/W
40022
SETPOINT 3 L
R/W
40023
HYSTERESIS 1 H
R/W
40024
HYSTERESIS 1 L
R/W
40025
HYSTERESIS 2 H
R/W
40026
HYSTERESIS 2 L
R/W
40027
HYSTERESIS 3 H
R/W
40028
HYSTERESIS 3 L
R/W
40029
INPUTS
-
R
40030
OUTPUTS
NO
R/W
40037
Test weight for calibration H
Use with command ‘101’ of the
COMMAND REGISTER
R/W
40038
Test weight for calibration L
R/W
40043
Weight value corresponding to ZERO of the analog
output H
Only after command ‘99’ of the
Command Register.
(Analog Models Only)
R/W
40044
Weight value corresponding to ZERO of the analog
output L
R/W
40045
Weight value corresponding to Full Scale of the analog
output H
R/W
40046
Weight value corresponding to Full Scale of the analog
output L
R/W
REGISTER
DESCRIPTION
Saving to EEPROM
ACCESS
Important