12 setpoints programming – Rice Lake Signal Conditioning Transmitter/Indicators User Manual
Page 34

30 SCT Weight Transmitter Operator’s Manual
2.12 Setpoints Programming
1. Press
to enter setpoints and hysteresis settings.
2. Press
or
until desired setpoint or hysteresis parameter is displayed, press
to enter.
3. Press
or
until desired value is displayed, press
to confirm.
4. Press
to exit setpoints and hysteresis settings.
These values are set to zero if the calibration is changed significantly (see Section 2.1.1 “Theoretical
Calibration” on page 13 and Section 2.1.5 “Weight (Span) Calibration (With Test Weights)” on page 15).
Parameter
Choices
Description
SEtP 1
SEtP 2
SEtP 3
0-Full Scale
0 *
Setpoint; relay switching occurs when the weight
exceed the value set in this parameter. The type of
switching is settable (see Section 2.10 “Outputs
And Inputs Configuration” on page 27).
HYStE 1
HYStE 2
HYStE 3
0-Full Scale
0 *
Hysteresis, value to be subtracted from the
setpoint to obtain contact switching for decreasing
weight. For example with a setpoint at 100 and
hysteresis at 10, the switching occurs at 90 for
decreasing weight.
* - indicates default value.