14 SCT Weight Transmitter Operator’s Manual
7. Press
or
until desired display division size is displayed, press
.
8. This completes the Theoretical Calibration, press
twice to exit set-up menu or continue to
By modifying the theoretical full scale, the sensitivity or divisions, the Weight (Span) Calibration is cancelled
and the Theoretical Calibration only is considered valid.
If the theoretical full scale and the recalculated full scale in Weight (Span) calibration (see Section 2.1.5 on
page 15) are equal, this means that the calibration currently in use is Theoretical Calibration; if they are
different, the calibration in use is the Weight (Span) Calibration based on test weights.
By modifying the theoretical full scale, the sensitivity or divisions and all the system’s parameters containing a
weight value will be set to default values (setpoints, hysteresis, etc.).
2.1.2 Maximum Capacity (NASS )
Maximum capacity (live load/product) that can be displayed. When the weight exceeds this value by 9 divisions
the following is displayed ‘
------
’, indicating overload. To disable this function, set to 0.
1. Press and hold
, then press
to enter set-up menu, CALIb will be displayed.
2. Press ,
FS-tEO is displayed.
3. Press
or
until NASS is displayed, press
.
4. Press
or
until desired capacity is displayed, press
.
5. Press
twice to exit set-up menu.
2.1.3 Zero Setting
Perform this procedure after having set the Theoretical calibration, see Section 2.1.1 on page 13.
This menu may also be accessed directly from the weight display, press and hold
for 3 seconds.
1. Press and hold
, then press
to enter set-up menu, CALIb will be displayed.
2. Press ,
FS-tEO is displayed.
3. Press
or
until ZErO is displayed, press
.
4. The weight value to be set to zero is displayed. In this phase all of the LEDs are flashing. Press
, the
weight is set to zero (the value is stored to the permanent memory).
5. Press
twice to exit set-up menu.
Press
to display the accumulated deadload that has been zeroed off by the instrument, displaying the sum
of all of the previous zero settings.
2.1.4 Zero Value Manual Entry
Perform this procedure only if it is not possible to zero off the scale structure, for example because it
contains product that can not be unloaded.
Enter the estimated structured dead load value that would be zeroed.
1. Press and hold
, then press
to enter set-up menu, CALIb will be displayed.
2. Press ,
FS-tEO is displayed.
3. Press
or
until INP 0 is displayed, press
.
4. Press
or
until desired dead load value is displayed, press
.
5. Press
twice to exit set-up menu.
Note
Note
Note
Important