5 program selection - reverse, 6 program selection - not legal, 7 keypad or display locking – Rice Lake Signal Conditioning Transmitter/Indicators User Manual
Page 36: 5program selection - reverse, 6program selection - not legal, 7keypad or display locking

32 SCT Weight Transmitter Operator’s Manual
2.13.5Program Selection - Reverse:
Scale capacity is displayed when scale is empty. As weight is added display will count down.
1. With power off, press and hold
, then power on. PrOG is displayed.
2. Press ,
bASE is displayed.
3. Press ,
rEuEr is displayed.
4. Press ,
UAIt is displayed.
5. Instrument will reboot.
2.13.6Program Selection - Not Legal:
Scale approval state is preset for not-legal. For other choices please contact technical assistance:
nOtLEG: Not Legal for Trade.
LEGAL: Legal for Trade (OIML)
NuLt: Legal for Trade, multi-interval (OIML)
1. With power off, press and hold
, then power on. PrOG is displayed.
2. Press ,
bASE is displayed.
3. Press ,
NOtLEG is displayed.
4. Press ,
UAIt is displayed.
5. Instrument will reboot.
By pressing
you will quit the program without introducing any changes and without deleting any of the set
variables.
If you do not have a specific manual for the newly set program, you can request it from technical assistance.
2.13.7Keypad Or Display Locking
1. Press
immediately followed by
, hold them down for about 5 seconds (this operation is also possible
via the MODBUS and ASCII protocols):
2. Press
or
until desired parameter is displayed, press
.
Parameter
Description
FrEE
no lock
KEY
keypad lock: if active, when key is pressed the message bLOC is
displayed.
dI SP
Keypad and Display lock: if active, the keypad is locked and the
display shows the instrument model (weight is not displayed); by
pressing a key the display shows bLOC for 3 seconds.
Note