Status register, Possible command to send to the command register, Real calibration commands (with sample weight) – Rice Lake Signal Conditioning Transmitter/Indicators User Manual
Page 44

40 SCT Weight Transmitter Operator’s Manual
Status Register
Bit 13 (writing error) rises if the PLC tries to write a value out of a parameter’s permitted interval. In this case the
writing has no effect and the previous parameter value is maintained.
Possible Command To Send To The Command Register
Commands are sent only once; to reassert it write the value 0 to the Command register and then the command
again
Real Calibration Commands (With Sample Weight)
The instrument calibration can be changed via PROFIBUS.
To correctly set the sample weight, consider the value of the division module (e); e.g. if you want to set the value to
100kg and the division value is 0.001, set the register value to 100000 (you remove the decimal points form the
value that would be 100.000 with decimals).
To perform the edit calibration procedure, it needs to unload the system and zero-setting the displayed weight value
with the command 100 “Zero-setting for calibration” of Command Register. Then, put a sample weight onto the
system, write the correct known weight value to the “Sample Weight for calibration” register; to save in memory
the sample weight value, send the command 101 “Sample weight storage for calibration”.
Bit 0
Cell Error
Bit 1
A/D Converter Malfunction
Bit 2
Maximum weight exceeded by 9 divisions
Bit 3
Gross weight higher than 110% of full scale
Bit 4
Gross weight beyond 999999 or less than -999999
Bit 5
Net weight beyond 999999 or less than -999999
Bit 6
Bit 7
Gross weight negative sign
Bit 8
Net weight negative sign
Bit 9
Peak weight negative sign
Bit 10
Net display mode
Bit 11
Weight stability
Bit 12
Weight within +/-¼ of a division around ZERO
Bit 13
Writing error
Bit 14
Bit 15
0
No command
17
Reserved
7
NET display
18
Reserved
8
SEMI-AUTOMATIC ZERO
21
Keypad lock
10
Reserved
22
Keypad and display unlock
11
Reserved
23
Keypad and display lock
12
Reserved
99
Save data in EEPROM
13
Reserved
100
Zero-setting for calibration
14
Reserved
101
Sample weight storage for
calibration
15
Reserved
9999
Reset (reserved)
16
Reserved
Note