0 operation, 1 semi-automatic tare (net/gross), 2 preset tare (subtractive tare device) – Rice Lake Signal Conditioning Transmitter/Indicators User Manual
Page 37: 4 peak, Operation

Operation 33
3.0 Operation
3.1
Semi-automatic Tare (Net/Gross)
The semi-automatic tare value is lost upon instrument power-off.
The semi-automatic tare operation is not allowed if the gross weight is zero.
1. To capture tare and weigh in net mode (SEMI-AUTOMATIC TARE), close the NET/GROSS input or
press
for 3 seconds. The instrument displays the net weight (zero) and the NET LED lights up.
2. To display the gross weight again, keep the NET/GROSS input closed or press
for 3 seconds.
3. This operation can be repeated by the operator to allow the loading of several products.
Press and hold
to display the gross weight temporarily. When
is released, the net weight will be
displayed again.
3.2
Preset Tare (Subtractive Tare Device)
It is possible to manually set a preset tare value to be subtracted from the display value provided that the P-tArE ≤
max capacity.
1. Press and hold
and
to display P-tArE, press
.
2. Press
or
until desired value is displayed, press
.
3. Press
to exit P-tArE.
4. After setting the tare value, go back to the weight display, the display shows the net weight (subtracting the
preset tare value) and the NET LED lights up to show that a tare has been entered.
Press and hold
for 3 seconds to display the gross weight temporarily. When
is released, the net weight
will be displayed again.
To delete a preset tare and return to the gross weight display:
1. Press hold
for 3 seconds or keep the NET/GROSS input (if any) closed for the same length of time (3
seconds). The preset tare value is set to zero. The NET LED is turned off when the gross weight is
displayed once again.
If a semi-automatic tare (net) is entered, it is not possible to access the enter preset tare function.
If a preset tare is entered, it is still possible to access the semiautomatic tare (net) function. The two different
types of tare are added.
All the semi-automatic tare (net) and preset tare functions will be lost when the instrument is turned off.
3.3
Semi-automatic Zero (Weight Zero-setting For Small Variations)
By closing the SEMI-AUTOMATIC ZERO input, the weight is set to zero. The zero setting will be lost when the
instrument is turned off.
This function is only allowed if the weight is lower than the 0 set value (see 0 SET in Section 2.3 on page 18),
otherwise the
t
-----
alarm appears and the weight is not set to zero.
3.4
Peak
By keeping the input closed the maximum weight value reached remains displayed. Opening the input the current
weight is displayed.
If you wish to use this input to view a sudden variation peak, set the FILTER ON THE WEIGHT to 0.
Note
Note
Note
Note
Note