General Tipping Bucket Problems

The following table contains general techniques for troubleshooting
the tipping bucket.

Warning: Contact your regional ADS representative for
further diagnosis before replacing a tipping bucket.

Problem

Monitor is not registering or recording tips.

Possible

Causes

Debris may be present in the funnel.

Cable connecting the tipping bucket to the monitor may
be loose or broken.

Tipping bucket may not be level.

Tipping bucket and/or tipping mechanism may be faulty.

Circuit board is not responding or may be faulty.

Possible

Solutions

Clean out the funnel in the tipping bucket.

Secure tipping bucket connection to monitor (if
necessary).

Check and adjust the level, and then secure the base
plate or tipping bucket screws.

Replace the tipping bucket.

Replace the monitor if defective.