beautypg.com

Safety device and error code – Yamato Scientific SQ 510 Autoclave, Sterilizers User Manual

Page 49

background image

46

In the Event of Failure…

Safety Device and Error Code

Internal communication
error

Er.17” appears

• Communication error between the control board and

display board.

™

Make a call for service.

Water level error

Er.20” appears

• Lack of water supply
™

Supply water.

™

Check the amount of water to be supplied referring to 9 of

"Pour water into the chamber" in Page 11. If the error is
not cancelled, contact our service department.

Safety valve

Safety valve is

operated.

• Pressure rise inside the chamber or safety valve

failure.

™

Make a call for service.