Before using this unit, Requirements for installation – Yamato Scientific SQ 510 Autoclave, Sterilizers User Manual
Page 11

8
Before Using This Unit
Requirements for Installation
7. Choose a correct power distribution board or receptacle
• Choose a correct power distribution board or receptacle that meets the unit’s rated electric
capacity.
• Operating voltage range for respective equipment models are as follows.
SN200/300/500 and SQ500 models: AC100 to 120V
SN210/310/510 and SQ510models: AC200 to 240V
Electric capacity:
SN200: AC100V-120V 12.5A-15A, SN210: AC200V-240V 6.5A-7.5A
SN300: AC100V-120V 16.5A-20A, SN310: AC200V-240V 8.5A-10A
SN500: AC100V-120V 19.5A-23.5A, SN510: AC200V-240V 10A-12A
SQ500: AC100V-120V 20.5A-24.5A, SQ510: AC200V-240V 10.5A-12.5A
NOTE)
There could be the case that the unit does not run even after turning ON the power. Inspect
whether the voltage of the main power is lowered than the specified value, or whether other
device(s) uses the same power line of this unit. If the phenomena might be found, change the
power line of this unit to the other power line.
• Starburst connection with a branching receptacle or extended wiring with a cord reel lowers
electrical power voltage, which may cause the degradation of refrigeration capability.
• Connect the unit to only the power supply. If it is connected to a gas pipe, water pipe or
telephone line, an accident or malfunction may result.
8. Handling of power code
• Do not entangle the power cord. This will cause overheating and possibly a fire.
• Do not bend or twist the power cord, or apply excessive tension to it. This may cause a fire
and electrical shock.
• Do not lay the power cord under a desk or chair, and do not allow it to be pinched in order to
prevent it from being damaged and to avoid a fire or electrical shock.
• Keep the power cord away from any heating equipment such as a room heater. The cord's
insulation may melt and cause a fire or electrical shock.
• If the power cord becomes damaged (wiring exposed, breakage, etc.), immediately turn off the
power at the rear of this unit and shut off the main supply power. Then contact your nearest
dealer for replacement of the power cord. Leaving it may cause a fire or electrical shock.
• Connect the power plug to the receptacle which is supplied appropriate power and voltage.