Before using this unit, Installation procedure – Yamato Scientific SQ 510 Autoclave, Sterilizers User Manual
Page 16

13
Before Using This Unit
Installation Procedure
16 Precautions for drainage
• Before draining the water, make sure that the pressure, equipment temperature and water
temperature inside the chamber have decreased sufficiently.
17 Precautions for continuous operation
• When operating the equipment continuously after sterilization is completed, leave the
equipment for about 15 minutes with the cover opened to sufficiently lower the temperature
inside the chamber and then close the cover. If the temperature is high, the cover may not
close due to high internal pressure of chamber.
• Before operating the equipment again, check the water level of drain bottle by observing it from
the observation window to make sure that it is below the level indicated by the drain level seal
on the door. In case the water level of drain bottle is above the drain level seal, drain the
water until the water level comes to 1500cc of the water level gauge, indicated on the side face
of bottle. The hot water or vapor may blow out from the drain bottle if the equipment is
operated with too much drain water (above the seal position).