Operation Method

Sterilization & Warm/Melting & Warm/Manual Course

MANUAL

MELT

PRE HEAT

AUTO

TEMP

PRES.

STERILIZE/MELT

HEAT

EXHAUST

WARM

SAMPLE

STERILIZE

WARM

/LIQUID

COOLING

FAN

WARM

END

TIME

MAINT.

ENTER

ENTER

WARM

STERILIZER
/MELT

STERILIZER
/MELT

START

WARM

STERILIZE

/NORMAL

STERILIZE

SET

MEMORY

START/

STOP

6. Set the heat-retention time

The current heat-retention time is displayed with blinking on

the time display screen. The STERILIZER/MELT lamp
blinks. Press the ∧∨ on the TIME key and set the desired
heat-retention time.

™

Set "00:00" when skipping the heat-retention process in the

manual operation.

After setting the time, press the ENTER key to determine the

value.

MANUAL

MELT

PRE HEAT

AUTO

TEMP

PRES.

STERILIZE/MELT

HEAT

EXHAUST

WARM

SAMPLE

STERILIZE

WARM

/LIQUID

COOLING

FAN

WARM

END

TIME

MAINT.

ENTER

ENTER

WARM

STERILIZER
/MELT

STERILIZER
/MELT

START

WARM

STERILIZE

/NORMAL

STERILIZE

SET

MEMORY

START/

STOP

7. Start operation

• Press the START/STOP key.

The equipment starts operation with the HEAT lamp blinks.

8. End operation

• The time display screen displays “END” when the operation is

completed.
Press the START/STOP key. The equipment goes into the
standby state. The lever on the cover is unlocked and the
cover can be opened.

To change or confirm the preset value during operation…

Preset values can not be changed while the equipment is in operation. Before starting operation, make
sure to confirm them and change them if required.
Press the course key currently being operated to confirm the preset value while the equipment is in
operation. The preset value has been displayed while the key is pressed.

To abort operation…

Press the START/STOP key. The equipment stops operation.
In case the operation is aborted before the pressurization process, the END lamp lights up and the time
display screen displays “END” without a buzzer sound.
In case the operation is aborted after the pressurization process, the equipment goes into the exhaust
process.

Power-off during operation

If the power is turned off due to a power failure, the operation is aborted. After the recovery, the “bLAc
oUt” is displayed to notify the power failure. Press any key to delete the “bLAc oUt”.
In case the operation is aborted before the pressurization process, the END lamp lights up and the time
display screen displays “END” without a buzzer sound.
In case the operation is aborted after the pressurization process, the equipment goes into the exhaust
process.
The operation continues if aborted during the heat-retention process.