Operation method, Auto start function, When starting the operation at 12:00 – Yamato Scientific SQ 510 Autoclave, Sterilizers User Manual
Page 39

36
Operation Method
Auto Start Function
Auto start function
This function automatically starts the operation of equipment at the specified time with the selected
course. The time can be set in increments of one minute within the range from 00:00 to 23:59.
The auto start function remains after the power is restored due to a power failure.
MANUAL
MELT
PRE HEAT
AUTO
TEMP
PRES.
STERILIZE/MELT
HEAT
EXHAUST
WARM
SAMPLE
STERILIZE
WARM
/LIQUID
COOLING
FAN
WARM
℃
END
TIME
MAINT.
ENTER
ENTER
WARM
STERILIZER
/MELT
STERILIZER
/MELT
START
WARM
STERILIZE
/NORMAL
STERILIZE
SET
MEMORY
START/
STOP
①
Make sure that the course to be operated is selected and then
press the AUTO START key. The AUTO START lamp lights up.
TIME
ENTER
WARM
STERILIZER
/MELT
②
The auto start time screen is displayed.
The time display screen displays the auto start time currently
set with blinking. Press the ∧∨ on the TIME key and set the
desired time.
When starting the operation at 12:00…
TIME
ENTER
WARM
STERILIZER
/MELT
START/
STOP
③
Set the desired auto start time and then press the ENTER key
or START/STOP key to determine the value.
AUTO
SAMPLE
TIME
ENTER
WARM
STERILIZER
/MELT
START
START/
STOP
④
The AUTO START lamp continues blinking until the equipment
starts operation.
Press the START/STOP key while the AUTO START lamp is
blinking to confirm the operation start time.
The procedure above is possible only when the cover on the
main unit is locked.
⑤
Press the AUTO START key to abort the preheating operation.