Before Using This Unit

Installation Procedure

Reference Data

Sterilizing operation using disposal bag for biochemically dangerous object

Open the opening of sterilization bag so the vapor can be easily entered into it.

Secure the bag with a wire rack so it should not fall down during operation.

The height of bag should be about two-thirds of chamber. If it is too high, the

vapor cannot be easily entered into it, or it blocks the vapor outlet at the upper
part of chamber, which may cause insufficient sterilization.

The preset temperature should be the upper temperature limit of bag to be used.

The preset time varies depending on the quality and quantity of sterile samples. Refer to the

following data for the preset time.

• Reference example of SN500 model at room temperature of 25℃

Sterile sample

Sterilization temperature Sterilization

time

Note

Gauze

121℃

30 min.

Five bolts of dry gauze

Petri dish

121℃

40 min.

30 Petri dish with a cover

™

The data above, however, is used as reference. The actual sterile condition varies depending on the

characteristics and quantity of samples or type of vessels to be used. Confirm the sterile condition
using the biological indicator or chemical indicator.

Sterilization

bag