Before Using This Unit
Installation Procedure
Reference Data
Sterilizing operation using disposal bag for biochemically dangerous object
①
Open the opening of sterilization bag so the vapor can be easily entered into it.
Secure the bag with a wire rack so it should not fall down during operation.
②
The height of bag should be about two-thirds of chamber. If it is too high, the
vapor cannot be easily entered into it, or it blocks the vapor outlet at the upper
part of chamber, which may cause insufficient sterilization.
③
The preset temperature should be the upper temperature limit of bag to be used.
④
The preset time varies depending on the quality and quantity of sterile samples. Refer to the
following data for the preset time.
• Reference example of SN500 model at room temperature of 25℃
Sterile sample
Sterilization temperature Sterilization
time
Note
Gauze
121℃
30 min.
Five bolts of dry gauze
Petri dish
121℃
40 min.
30 Petri dish with a cover
The data above, however, is used as reference. The actual sterile condition varies depending on the
characteristics and quantity of samples or type of vessels to be used. Confirm the sterile condition
using the biological indicator or chemical indicator.
Sterilization
bag