12

Before Using This Unit

Installation Procedure

11 Set the sterile samples

• Set the samples to the chamber, putting them into the attached rack or cast (sold separately).
• Put the sample or sterilization bag into the chamber so they should not block or cover the

sensor inside the chamber, exhaust outlet and end connection to pressure gauge. If they are
blocked or covered, the vapor cannot be discharged and the equipment cannot be operated
correctly. Do not spill the samples when taking them out from/putting them into the chamber.
The failure in piping system, bad smell or dirt may result in.

• In case liquid such as medicinal solution or medium is sterilized, the amount of liquid should be

60% or less of the capacity of container. They may be boiled over if too much quantity is
supplied.

• Widely open the opening of sterilization bag when used. If it is closed, the samples are

insufficiently sterilized.

12 Close the door before operation

• Make sure to close the door of equipment before operation. If the door is not fully closed, the

drain bottle falls down and may cause a burn injury. Do not open the door during operation.

Close the cover

• Make sure that no foreign objects exist on the

packing of the cover and its contact area before
closing the cover. If any foreign object exists,
the vapor may leak from the inside.

Turn the lock lever to the left to close the cover.

Turn the lock lever to the right to open the cover.

13

• The lock lever on the cover is held by the hook and does not move when the cover remains

open.

• Fully close the cover and slide the lock lever on the cover to the left side. If it is closed

inappropriately, the vapor blows out from the inside, which may cause a burn injury.

• Do not press the hook and operate the lock lever for purposes other than maintenance of

equipment.

14 Attach the vapor cup

• The equipment is equipped with a vapor cup to

prevent the vapor that generate during air purge
from becoming water droplet and dropping down
onto the floor. Attach the cup onto the vapor
outlet on the left side face of equipment referring
to the right figure. Make sure to pour off the
water inside the vapor cup after every operation.

15 Attach the droplet tray

• Attach the droplet tray to the equipment to

prevent the waterdrops that are made from vapor
that generate during air purge from dropping
down from the packing onto the samples or onto
the top board. Pour off the droplets inside the
tray periodically.

Vapor cup

Droplet tray