Operation Method

Forced Cooling Function / Preheating Function

Forced cooling function

This function turns on the cooling fan during exhaust process to shorten the cooling time.
The cooling fan is turned on during switching to the exhaust process in the apparatus sterilization
course. In other courses, it starts to run at the saturated vapor temperature (refer to the graph on
Page 15) of -2℃ or less. It stops when the equipment goes into the standby state after operation is
completed, or when the temperature inside the chamber reaches 60℃. The function can be set
anytime before and during operation of equipment.

MANUAL

MELT

PRE HEAT

AUTO

TEMP

PRES.

STERILIZE/MELT

HEAT

EXHAUST

WARM

SAMPLE

STERILIZE

WARM

/LIQUID

COOLING

FAN

WARM

END

TIME

MAINT.

ENTER

ENTER

WARM

STERILIZER
/MELT

STERILIZER
/MELT

START

WARM

STERILIZE

/NORMAL

STERILIZE

SET

MEMORY

START/
STOP

Press the COOLING FAN key.

The COOLING FAN lamp lights up and the function becomes
available.

Preheating function

This function keeps the temperature of feedwater inside the chamber with the preset temperature.
The range of preset temperature is from 45℃ to 80℃. The operation automatically ends after five
hours.
The preheating operation function remains after the power is restored due to a power failure.

MANUAL

MELT

PRE HEAT

AUTO

TEMP

PRES.

STERILIZE/MELT

HEAT

EXHAUST

WARM

SAMPLE

STERILIZE

WARM

/LIQUID

COOLING

FAN

WARM

END

TIME

MAINT.

ENTER

ENTER

WARM

STERILIZER

/MELT

STERILIZER

/MELT

START

WARM

STERILIZE

/NORMAL

STERILIZE

SET

MEMORY

START/
STOP

Press the PRE HEAT key.

ENTER

ENTER

TEMP

WARM

TIME

WARM

STERILIZER

/MELT

STERILIZER

/MELT

The preheating function screen is displayed.

The time display screen displays the character "PrEH", which
indicates the preheating function. The temperature display
screen displays the value (temperature). The temperature
display screen displays with blinking the preheating
temperature currently set and the STERILIZE/MELT lamp
blinks.
Press the ∧∨ on the TEMP key to set the desired preheating
temperature.

MANUAL

MELT

PRE HEAT

AUTO

TEMP

PRES.

STERILIZE/MELT

HEAT

EXHAUST

WARM

SAMPLE

STERILIZE

WARM

/LIQUID

COOLING

FAN

WARM

END

TIME

MAINT.

ENTER

ENTER

WARM

STERILIZER

/MELT

STERILIZER

/MELT

START

WARM

STERILIZE

/NORMAL

STERILIZE

SET

MEMORY

START/
STOP

After setting the temperature, press the ENTER key or

START/STOP key to determine the value. This starts the
preheating operation.

Press the PRE HEAT key to abort the preheating operation.