Before Using This Unit
Installation Procedure
7 Set the attached drainboard onto the bottom
surface inside the chamber
• The drainboard stabilizes the sterile samples
inside the chamber as well as protects the
heater and sensor. Make sure to set it.
8 Close the drain valve
• Close the drain valve at the bottom on the left side face of main
unit. Water leak occurs if not fully closed, which may cause
the burn injury or no-load (water) operation.
• Connect an appropriate hose (inner diameter: 12mm) to the
drain valve and lead it to the draining site.
9 Pour water into the chamber
• Before setting the sterile samples, pour water into the
chamber to the water level gauge (notch) position.
• Insufficient water may cause the no-load (water)
operation. Check the water level every time before
operation. Refill it before the level becomes too low.
Water is required to be poured at dissolution
operation, as well as sterilizing operation.
• When the water level lowers, the equipment detects
an abnormality (“Err20”) and cuts off the heater.
Depending on the conditions of equipment, however,
the detection requires too much time, which may
cause the heater deterioration. Refill water before
the water level becomes too low.
• Refer to the right table for the quantity of water to be
refilled.
Water QTY for models
SN200/210
2200~2300 ㏄
SN300/310
3500~3700 ㏄
SN500/510
3500~3700 ㏄
SQ500/510
6300~6500 ㏄
10 Use distillated or purified water for sterilizing water
• Fill distillated or purified water inside the chamber. Tap water may be used. Calculus
generate inside the chamber when tap water is used. Frequent cleaning is therefore required.
• Do not use well water. It may cause the corrosion or dirt inside the chamber.
Set the drainboard
horizontally.
Close
Open
Water level
gauge
High
Low