Operation method, Operation course/function list – Yamato Scientific SQ 510 Autoclave, Sterilizers User Manual

Page 26

background image

23

Operation Method

Operation Course/Function List

Name Description

Page

This course sterilizes and warms other samples other than the above.

Manual course

• Press the MANUAL key to go into the operation setting mode.If other

course has been selected, press the MANUAL key again.

• Set the sterilization time and sterilization temperature.
• Set the temperature with the TEMP key and then press the ENTER key.
• Set the time with the TIME key and then press the ENTER key.
• Set the heat-retention time and heat-retention temperature.
• Set the temperature with the TEMP key and then press the ENTER key.
• Set the time with the TIME key and then press the ENTER key.
• Press the START/STOP key to start operation.
• Press the START/STOP key again to stop operation.

29

™

The current operation mode can not be changed during operation of equipment. To change the

operation mode, stop the operation of equipment.


The equipment has the following functions.

Name Description

Page

Calendar setting

This function is included in the maintenance mode.
It sets the dominical year, month, date and time.

Keylock function

This function is included in the maintenance mode.
It disables all key operations, except the START/STOP key operation
and cancellation of keylock state.
The “_Loc” is displayed if an unavailable key operation is done. (※)

Pattern lock function

This function is included in the maintenance mode.
It disables the change related to operation course and memory.
The “PLoc” is displayed if an unavailable key operation is done. (※)

Buzzer function

This function is included in the maintenance mode.
It mutes the key operation sound except for the buzzer sounds at
warning and operation end.

Error log display

This function is included in the maintenance mode.
It displays up to 20 errors occurred in the past, including the error
content and time of occurrence.

Setting of sample
temperature sensor

This function is included in the maintenance mode.
It enables the sample temperature function. If the setting is turned to
ON when the optional sample sensor is not attached, the “Er.8”, which
indicates disconnection of sample sensor, occurs.

Accumulated time

This function is included in the maintenance mode.
It displays the accumulated current-carrying time by the hour.

31

: When both of keylock and pattern lock functions have been set, "_Loc" is displayed by priority.