[ Care and Use ManUal ]

Glyco-Pak DEAE Columns

6

To aid in diagnosing potential column problems:

Equilibrate the column in 15 mM sodium acetate adjusted to a pH of 5.0
with acetic acid at a flow rate of 0.8 ml/min for 1 and 1/2 hours

Inject 10 µl of a 10 mg/ml solution of N-acetylneuraminic acid in
15 mlVl sodium acetate buffer at pH 5.0

Set detector to 220 nm.

Retention time will vary with absolute concentration of the buffer as well as
instrument configuration, however, the peak shape and retention should be
similar to that seen in the chromatogram below.

Figure 8: Test Chromatogram, N-Acetylneuraminic Acid

Performing this test prior to discussing the problem with your Waters repre-
sentative will help in diagnosing the problem.

a. Column Storage

Store columns in distilled water and refrigerate at 4 - 6 °C. DO NOT
FREEZE.

DO NOT allow buffers or other potentially harmful materials to remain
in the system overnight when not being used. Run buffers with a slow
flow rate (0.1 ml/min) overnight if necessary.

Return the column to its box with the end caps firmly in place for stor-
age. Allowing columns to dry out can result in poor chromatographic
performance.

b. Troubleshooting

Use Table 2 to troubleshoot problems with your column.

Table 2: Column Problems and Solutions

V. warranty and serVIce InformatIon

a. Service Information

Waters Corporation staff of experienced service specialists provide mainte-
nance assistance on both preventative and/or corrective levels. For complete
information and assistance, please call Waters Service Department at 1-800-
252-HPLC. For solutions to particular applications questions Waters team of
technical support personnel are available to help you with specialized support.
They may be contacted at 1-800-252-HPLC in Milford, MA.

Waters Corporation staff of experienced service specialists provide mainte-
nance assistance on both preventative and/or corrective levels. For complete
information and assistance, please call Waters Service Department at 1-800-
252-HPLC. For solutions to particular applications questions Waters team of
technical support personnel are available to help you with specialized support.
They may be contacted at 1-800-252-HPLC in Milford.

Problem

Cause

Solution

Excess pressure

buildup

• Filters plugged with

Particulates

• Sample precipitates

on column (sample

not soluble in

mobile phase)

• Clean in an ultrasonic bath or

replace.

• Always alter mobile phases

and samples.

• Slowly purge with a strong

mobile phase that is both

appropriate to dissolve the

contaminate and compatible

with the column.

Loss of resolution,

broad peaks, low

plate counts

• Filters partially

plugged

• Replace or clean inlet and

outlet filters in an ultrasonic

bath.