Care and use manual – Waters End Connector Kit User Manual

Page 2

background image

[ Care and Use ManUal ]

2

IV. USIng the cart rIdge

Use the cartridge as you would any other analytical HPLC column. Be
aware of the flow direction arrow printed on the label of the cartridge.
Use a precolumn in-line filter to prevent cartridge filter blockage.
Please refer to the Waters Analytical Columns and Cartridges Care
and Use Instructions Sheet for mobile phase and sample guidelines,
operating parameters, solvent-viscosity tables, chromatography
guidelines, and other useful information.

V. replacIng the cart rIdge

No tools are required to replace the cartridge.

1. Unscrew the endfittings and remove the cartridge. It is not
necessary to disconnect the endfittings from the inlet and outlet
tubing.

2. Remove the c-clips and the retaining nuts from the old cartridge.
Discard the cartridge according to appropriate regulations.

3. Install a new cartridge (see Section III).

VI. replacIng the o-rIngS In the endfIt tIngS

To replace the o-ring, you will need a replacement o-ring kit (2/pkg),
part number WAT023401; a small, flat-blade screwdriver; and a sharp
needle.

Figure 2: Replacing the O-Ring

To replace the o-ring:

1. Unscrew the endfitting assembly.

2. Remove the white plastic cover.

3. Remove the e-ring from the connector using a screwdriver.

4. Take the connector out of the connector housing. Use a sharp
needle to remove the old o-ring from the connector.

Caution: Do not scratch the groove in the connector. This may damage
the sealing surface.

5. Check the groove and the new o-ring. Remove any particulate
matter. Insert the new O-ring into the groove.

6. Reassemble.

VII. t roUbleShootIng

General care and maintenance is addressed in the Waters Analytical
Columns and Cartridges Care and Use Instruction Sheet. Additional
symptoms you may see when using the steel cartridge include:

endfitting assembly

groove

retaining nut

cartridge

c-clip

Figure 1: Assembling the cartridge column

connector

o-ring

plastic cover

connector housing

e-ring

Symptom

cause

corrective action

Solvent leaks from

connector

Cartridge

not seated in

endfitting

Disassemble, properly

seat cartridge

Particles on

the FACE of the

o-ring

Wipe clean with a cotton

tip applicator

Worn o-ring

Replace the o-ring

Broad peaks, loss of

peak resolution

Packing material

fouled

Replace the new cartridge

column, reuse endfittings

and c-clips (see Section V).

High backpressure

Filters in car-

tridge plugged