[ Care and Use ManUal ]
2
IV. USIng the cart rIdge
Use the cartridge as you would any other analytical HPLC column. Be
aware of the flow direction arrow printed on the label of the cartridge.
Use a precolumn in-line filter to prevent cartridge filter blockage.
Please refer to the Waters Analytical Columns and Cartridges Care
and Use Instructions Sheet for mobile phase and sample guidelines,
operating parameters, solvent-viscosity tables, chromatography
guidelines, and other useful information.
V. replacIng the cart rIdge
No tools are required to replace the cartridge.
1. Unscrew the endfittings and remove the cartridge. It is not
necessary to disconnect the endfittings from the inlet and outlet
tubing.
2. Remove the c-clips and the retaining nuts from the old cartridge.
Discard the cartridge according to appropriate regulations.
3. Install a new cartridge (see Section III).
VI. replacIng the o-rIngS In the endfIt tIngS
To replace the o-ring, you will need a replacement o-ring kit (2/pkg),
part number WAT023401; a small, flat-blade screwdriver; and a sharp
needle.
Figure 2: Replacing the O-Ring
To replace the o-ring:
1. Unscrew the endfitting assembly.
2. Remove the white plastic cover.
3. Remove the e-ring from the connector using a screwdriver.
4. Take the connector out of the connector housing. Use a sharp
needle to remove the old o-ring from the connector.
Caution: Do not scratch the groove in the connector. This may damage
the sealing surface.
5. Check the groove and the new o-ring. Remove any particulate
matter. Insert the new O-ring into the groove.
6. Reassemble.
VII. t roUbleShootIng
General care and maintenance is addressed in the Waters Analytical
Columns and Cartridges Care and Use Instruction Sheet. Additional
symptoms you may see when using the steel cartridge include:
endfitting assembly
groove
retaining nut
cartridge
c-clip
Figure 1: Assembling the cartridge column
connector
o-ring
plastic cover
connector housing
e-ring
Symptom
cause
corrective action
Solvent leaks from
connector
Cartridge
not seated in
endfitting
Disassemble, properly
seat cartridge
Particles on
the FACE of the
o-ring
Wipe clean with a cotton
tip applicator
Worn o-ring
Replace the o-ring
Broad peaks, loss of
peak resolution
Packing material
fouled
Replace the new cartridge
column, reuse endfittings
and c-clips (see Section V).
High backpressure
Filters in car-
tridge plugged