Care and use manual – Waters DisQuE Dispersive Sample Preparation Kit User Manual
Page 3

D
is
Q
u
E QuEChERS Sample Preparation Products and Kits
3
[ Care and Use ManUal ]
Tip: For some commodities that contain a high water content, it is 
recommended that the extraction solvent is added to the 50 mL 
extraction tubes prior to the addition of the sample. This prevents 
undesirable caking of the anhydrous magnesium sulfate to the sample.
3. Dispersive Solid-Phase Extraction
A 1 mL aliquot of the top layer extract from the 50 mL extraction 
tube is transferred to the 2 mL DisQuE clean-up tube. Shake the tube 
vigorously and invert the tube a few times to ensure the extract is 
thoroughly mixing with all the solids in the 2 mL tube. The sample 
can be analyzed directly or diluted appropriately with a suitable 
buffer or solvent.
v. usIng tHe AoAC KIt And MetHod
The following section describes the general steps involved for a using 
DisQuE QuEChERS sample preparation kits for the AOAC (Association 
of Analytical Communities) procedure.
Extraction solution preparation
A solution of 1% acetic acid in acetonitrile is prepared by adding 
10 mL of glacial acetic acid to 500 mL of acetonitrile in a 
1 L volumetric flask. Bring to final volume with acetonitrile. Mix 
thoroughly. 
1. Sample Comminution
The sample is chopped into 2–5 cm
3
portions with a knife, and then
pulverized by a food blender. Blend until homogeneous.
2. Extraction/Partitioning
Add 15 mL of 1% acetic acid followed by the contents of the AOAC 
salt pouch to an empty 50 mL extraction tube. Manually shake a few 
times to ensure that no solids are stuck to the bottom of the tube. 
Transfer 15 g of the homogenized sub sample into the extraction 
tube. Shake vigorously for 1 minute. Centrifuge the tube at > 1500 
rcf for at least 1 minute. Caution: exceeding 10,000 rcf may dam-
age, deform, or break the extraction tube. 
3. Dispersive Solid-Phase Extraction
Transfer 1 mL of top layer extract from Tube 1 into the 2 mL DisQuE 
Clean-Up Tube 2. Shake vigorously for at least 30 seconds ensuring 
the sample and sorbents are thoroughly mixed. Centrifuge Tube 2 
at > 1500 rcf for 1 minute. Caution: exceeding 17,000 rcf may 
damage, deform, or break the centrifuge tube. Transfer the top 
extract into and autosampler vial for analysis. 
vI. stePs oF sAMPle CoMInutIon And solId-PHAse
eXtrACtIon usIng Cen KIt And MetHod
The following section describes the general steps involved for a using 
DisQuE QuEChERS sample preparation kits for the CEN procedure.
1. Sample Comminution
The sample is chopped into 2–5 cm
3
portions with a knife, and then
pulverized by a food blender. Blend until homogeneous.
2. Extraction/Partitioning
Add 10 mL of acetonitrile and the contents of the CEN method salt 
pouch to a 50 mL extraction tube. Manually shake a few times to 
ensure that no solids are stuck to the bottom of the tube. Transfer 
10 g of the homogenized sub sample into the extraction tube. Shake 
vigorously for 1 minute. Centrifuge the tube at > 1500 rcf for at 
least 1 minute. Caution: exceeding 10,000 rcf may damage, deform, 
or break the extraction tube.
3. Dispersive Solid-Phase Extraction
Transfer 1 mL of top layer extract from Tube 1 into the 2 mL DisQuE 
Clean-Up Tube 2. Shake vigorously for at least 30 seconds ensuring 
the sample and sorbents are thoroughly mixed. Centrifuge Tube 2 at 
> 3000 rcf for 5 minutes. Transfer the top extract into and autosam-
pler vial for analysis. Caution: exceeding 17,000 rcf may damage, 
deform, or break the centrifuge tube.
vII. storAge And dIsPosAl oF used CArtrIdges
The devices remain stable when stored in their original sealed pack-
aging. To store unused devices in opened pouches, squeeze the air 
out of the pouch, fold over the open end of the pouch twice, seal with 
tape, and store in a desiccator.
Dispose of used devices safely in accordance with applicable govern-
ment or local regulations.
vIII. AddItIonAl InForMAtIon
For more detailed information, consult the Waters Quality Parts
®
,
Chromatography Columns and Supplies Catalog by visiting 
www.waters.com/catalog, or contact your Waters representative.
If you require technical assistance, or have any questions 
regarding sample preparation, visit www.waters.com or call us at 
1-800-252-4752.