beautypg.com

Care and use manual – Waters GlycoWorks Single Use Sample Preparation Kit User Manual

Page 5

background image

GlycoWorks High-throughput Sample Preparation Kit

5

[ CARE AND USE MANUAL ]

Step 4. Removal of Excess Label

1) Condition a NEW well of the GlycoWorks cartridge by adding

1,000 μL of Milli-Q water to each cartridge and aspirate using the
vacuum manifold.

2) Add 1,000 μL of 85% acetonitrile to each well and aspirate using

the vacuum manifold or positive pressure manifold.

3) Add 200 μL of acetonitrile to the 20 μL of 2AB labeled glycan

sample and load into the cartridge using low vacuum setting
(1-2 in Hg).

4) Wash each well with 400 μL of 85% acetonitrile. Remove

filtrate to waste.

5) Replace the waste tray with sterile 1.5 mL Eppendorf tubes.

6) Elute glycans with 100 μL of 100 mM ammonium acetate in 5%

acetonitrile.

Note: The concentration of ammonium acetate and acetonitrile in the elu-
tion buffer may require optimization for some glycan samples.

7) Repeat elution two more times.

8) Place 96-well collection plate in a centrifugal vacuum evaporator

and dry the glycans to a concentrate of ≤ 2 µL in volume.

Note: If a rotor is not available for plates, transfer each eluent to a sterile
1.5 (or 0.5) mL Eppendorf and place in the vacuum evaporator. This may
contribute to a small sample loss due to extra transfer.

9) Glycans can be stored in Milli-Q water at -20 °C until required.

V. PREPARING LABELED GLYCANS FOR HILIC-FLR WITH
AN ACQUITY UPLC BEH GLYCAN COLUMN

The following steps are suggested for preparing the labeled glycans for
HILIC-FLR with an ACQUITY UPLC BEH Glycan column when using this
kit with the Waters Total Glycan Application Solution.

1. Reconstitute the labeled glycans in 100 µL of Milli-Q water by

gently aspirating, avoiding vortexing and sonication.

2. Add 150 µL of ACN to the reconstituted glycans immediately

prior to and in preparation for analysis by HILIC-FLR.

3. Inject 4 µL of the labeled glycan containing 60% acetonitrile

solution onto the column.

Note: Preparations of non-IgG glycoproteins or glycoprotein samples with
concentrations other than 1 mg/mL may need to be reconstituted with
different volumes to achieve desired results. Similarly, more concentrated
samples can be obtained by reconstituting with up to 10 times lower volumes.
The optimum injection volume for this application is ≤4 µL.

VI. REFERENCES

1. Lauber MA, Koza S, Fountain KJ. Optimization of HILIC SPE for the Quantita-

tive and Robust Recovery of N-Linked Glycans. Waters Application Note
720004710EN. 2013 June.

2. Lauber MA, Koza S, Fountain KJ.Single-Use and High-Throughput HILIC SPE

Device Formats and an IgG Control Standard for Facilitating N-Glycan Analy-
ses Waters Technology Brief 720004711EN. 2013 June.

3. Training Video: http://www.waters.com/waters/en_US/Glycan-Standards/nav.

htm?cid=134640534

Step 4 Tips and Tricks

Extended periods of time between incubation and analysis

may result in desialylation of labeled glycans and consequently
should be avoided.