II. STORAGE AND STABILIT Y

The standard mix comes in a Waters Crimp Top vial for direct injection
and contains 500 µL total in 1:10 in ACN:LC/MS Grade Water.

The standard mix is shipped at ambient temperatures. It is
highly recommended that upon receipt, the standard should be
refrigerated at 4 °C for short term and for long term stored frozen.

III. EXAMPLES OF USING T HE LCMS QC RM

In this section, two example methods are discussed to allow the
user to get an idea and understanding of how this standard can
be used under different conditions. The user is not aiming for an
exact replicate of these LC/MS chromatograms, however this is
provided as guidance. The advantage of this standard is that it can
be used with any column under different LC conditions. Therefore,
based on the information shown below, the user can customize this
standard to their specific method.

Shown in Figure 2 is a representative MS Spectrum under the
following conditions listed in Table 2.:

Table 2: Gradient Table for Figure 1

Time/min

Flow rate/

mL min

%A

%B

Curve

0.0

0.5

100

0

3.5

0.5

0

100

6

4.5

0.5

100

0

11

Total analysis time = 5 min

Column: ACQUITY

®

HSS T3, 2.1 x 100 mm, 1.8 µm

Mobile phase A:

Water + 0.1 % formic acid

Mobile phase B:

Acetonitrile

Injection volume: 1 µL

Acetaminophen

Sulfaguanidine

Sulfadimethoxine

Val -Tyr -Val

Verapamil

Terfenadine

Leucine -Enkephalin

Reserpine

Caffeine

Acetaminophen

Caffeine

Sulfaguanidine

Sulfadimethoxine

Val -Tyr -Val

Verapamil

Terfenadine

Leucine-Enkephalin

Reserpine

Table 3: Gradient Table for Figure 2

Time/min

Flow rate/

mL min

%A

%B

Curve

0.0

0.5

100

0

3.5

0.5

0

100

6

4.5

0.5

100

0

11

Total analysis time = 5 min

Figure 1: TOF MS ES+ Spectrum of the LCMS QCRM Standard using ACN Mobile Phase
with a SYNAPT

®

G2-S MS operated in positive ion mode in high resolution mode

Figure 2: TOF MS ES+ Spectrum of the LCMS QCRM Standard using Methanol
Mobile Phase with a SYNAPT G2-S MS operated in positive ion mode in high
resolution mode

Column:

ACQUITY HSS T3, 2.1 x 100 mm, 1.8 µm

Mobile phase A: Water + 0.1 % Formic acid
Mobile phase B:

Methanol + 0.1 % Formic acid

Injection volume: 1 µL