beautypg.com

Care and use manual, Iii. column use, A. sample preparation – Waters XBridge BEH Glycan, 2.5 μm XP and 3.5 μm Columns and Standards User Manual

Page 4: B. operating ph limits, Additive/buffer pk

background image

LC Systems with different system volumes would affect the final
recorded results.






















Figure 2. Comparative scaled separations (based on column particle size
differences) using Waters Glycan Performance Test Standard (P/N 186006349)
on ACQUITY H-Class UPLC System with ACQUITY BEH Glycan, 1.7 μm, XBridge
Glycan 2.5 μm XP, and XBridge Glycan 3.5 μm XP columns.

III. COLUMN USE

To ensure the continued high performance of XBridge BEH Glycan
Columns, observe the following guidelines:

a. Sample Preparation

1. Sample impurities often contribute to column contamination.

Samples should be free of particles before injection into
the system.

2. In most separations it is preferable to prepare the sample

in the gradient initial composition. However, the 2-AB
labeled glycans are often insoluble in the high acetonitrile
concentrations which typify HILIC initial conditions. Since
small volume injections are being made, the sample diluents
may contain higher aqueous content (e.g. 50%) than the initial
composition.

3. If the sample is not dissolved in the mobile-phase or solvent

combinations specified in this manual, ensure that the sample,
solvent, and mobile phases are miscible in order to avoid
sample and/or buffer precipitation. Preparation of 2-AB
labeled glycans involves one or two steps of solid-phase
extraction. As a result, protein precipitate has typically been
removed. If not, remove protein particles by centrifugation at
>10,000 rpm for more than 2 minutes.

b. Operating pH Limits

The recommended operating pH range for the XBridge BEH
Glycan Column is 3 to 8. A listing of commonly used buffers and
additives is given in Table 2. Additionally, the column lifetime
will vary depending on the operating temperature as well as the
type and concentration of buffer used.

Table 2: Buffer Recommendations for Using XBridge BEH Glycan Columns from pH 3 to 8

Additive/Buffer

pK

a

Buffer Range

(±1 pH unit)

Volatility

Used for

Mass Spec Comments

Acetic Acid

4.76

Volatile

Yes

Maximum buffering obtained when used with ammonium
acetate salt. Used in 0.1–1.0% range.

Formic Acid

3.75

Volatile

Yes

Maximum buffering obtained when used with ammonium
formate salt. Used in 0.1–1.0% range.

Ammonium (Acetate) 9.20

8.2–10.2

Volatile

Yes

Up to 100 mM.

Ammonium
(Formate)

9.20

8.2–10.2

Volatile

Yes

Up to 250 mM.

Triethylamine
(as acetate salt)

10.70

9.7–11.7

Volatile

Yes

Used in the 0.1–1.0% range. Volatile only when titrated with
acetic acid (not hydrochloric or phosphoric). Used as ion-pair
for DNA analysis at pH 7–9.

15 16

ACQUITY BEH Glycan 1.7 µm
2.1 x 150 mm

0.50 mL/min

XBRIDGE BEH Glycan 2.5 µm

XP

2.1 x 150 mm

0.34 mL/min

7.4

5.0

XBRIDGE BEH Glycan 3.5 µm
2.1 x 150 mm

0.24 mL/min

10.3

1 2

3

4

5

6

7

89

10

11

12

13

14

8700 psi (Column, Max)

3300 psi (Column, Max)

900 psi (Column, Max)

30.0 min

44.3 min

62.1 min

N-Acetylglucosamine

Manose

Galactose

Fucose

Sialic Acid

1

2

3

4

5

6,7

8

9

11

12

13

14

10

4

[ CARE AND USE MANUAL ]

XBridge BEH Glycan, 2.5 µm XP and 3.5 µm Columns and Standards